Outgoing Scsi Frame Settings - LeCroy Sierra M6-2 User Manual

Sas/sata protocol analyzer
Table of Contents

Advertisement

LeCroy Corporation

Outgoing SCSI Frame Settings

Before selecting Outgoing Frame Settings, select the Types Of Frames and the
Frame Number on which to inject an error. Then check Outgoing Frame in the Generate
Error On area and then the Settings button to display the Outgoing Frame Setting dialog.
Figure 2.135: SAS: Outgoing Frame Setting Dialog SCSI
Frame Type Error
Check this to introduce a frame type error.
Remove Frame from sequence
Check this to remove frame from sequence.
Delay in sending frame
Check this to delay sending frame and enter a value for the delay.
Frame Length Error
Check Frame Length Error, then choose the type of error to introduce.
Code violation and Disparity error
Check these and specify DWORD offset and the number of error DWORDs.
Don't check credit
Check this to disable credit checking.
SOF, EOF and CRC errors
Check these and specify Double or Omit by checking the corresponding option button.
136
 
S ierra M6‐2 SAS/SATA Protocol Analyzer User Manual
 