Adjustments and Replacements

MOTOR, NIP POSITION DRIVE (M2)

Prerequisites:
To Remove:
To Install:
3-54
Remove:
"BACK PANEL"
"RIGHT PANEL"
4 CIRCUIT BREAKER
CONNECTORS
4 SCREWS
BELT
TENSIONER
SCREW
1. Disconnect:
MOTOR CONNECTOR
4 CIRCUIT BREAKER CONNECTORS
2. Loosen the TENSIONER SCREW.
3. Disengage the BELT for the PULLEY on the NIP POSITION MOTOR.
4. Remove:
4 SCREWS
NIP POSITION MOTOR
1. Reverse the steps in the removal procedure.
MOTOR
CONNECTOR
NIP POSITION
MOTOR
Back of IMAGER