Installation Instructions
2-10
8. Install the CHARCOAL FILTER.
9. Close:
FRONT PANEL
TOP COVER
10. Open the package of electrical accessories.
IMPORTANT
:
The 5800 IMAGER will automatically detect the AC power line voltage and
will automatically configure itself for the applied voltage.
11. Select a POWER CORD.
Part number 7E9349 - for North American installations
Part number 7F6848 - for European installations
Part number 7F3990 - for Chinese installations
Check that the POWER CORD you select matches the AC power
OUTLETS at the site.
If the IMAGER is located outside the areas above, you can use one of
these POWER CORDS if the PLUG matches the power OUTLETS in the
building. If the customer cannot use any of the POWER CORDS
provided, the customer must obtain a "suitable" POWER CORD locally.
The customer must use an agency-approved POWER CORD with a
PLUG type "suitable" for the customer's building.