Kodak DryView 5800 Service Manual page 45

Laser imager
Hide thumbs Also See for DryView 5800:
Table of Contents

Advertisement

2G0947 - 31JAN08
SLOT
PROBE
4. Insert the PROBE into the slot on the right back of the PROCESSOR AY
and let the rotation of the PROCESSOR DRUM pull the PROBE until the
BLOCK reaches the slot.
5. Allow the temperature reading on the METER to stabilize. This requires
approximately 90 seconds.
6. Check that the temperature at the PROCESSOR DRUM is between
123.5° - 124.5° C (254.3° - 256° F).
7. Pull back slightly, 1 - 2 cm (0.4 - 0.8 in.), on the BLOCK to confirm that
the highest temperature is being read.
Adjustments and Replacements
PROCESSOR AY
DRUM
3-3