Kodak DryView 5800 Service Manual page 230

Laser imager
Hide thumbs Also See for DryView 5800:
Table of Contents

Advertisement

Diagnostics
Condition Code: 26-543
Web Portal Message
Film Jam in Area 3
5-82
Log Message
Film Jam Prior to Densitometer:
Film at Densitometer sensor
(S11) failed to detect film
leading edge.
Description
This error occurs if the leading edge is not detected at the FILM AT
DENSITOMETER SENSOR (S11) after the film leaves the exposure area. The
error is cleared when the TOP COVER is opened then closed. When this error
occurs, the PROCESSOR stops rotating and the HEATER is turned off. All
films in the printer will be marked bad and reprinted automatically.
System Impact
The printer is unable to print films. The spooler is still able to accept and queue
jobs but new films will not be initiated. All other films will be reprinted.
User Action
1. Open the IMAGER TOP COVER and remove any films in Area 3.
2. If no films were found, unlatch and open the PROCESSOR COVER.
3. Carefully remove any films found in the PROCESSOR.
4. Close and latch the PROCESSOR COVER.
5. Close the TOP COVER. If film was not previously removed, the
misplaced films will emerge from the IMAGER within 2 minutes.
FE Action
1. Clear the jam (see
page
5-111).
2. If the above action does not correct the problem, or if the problem recurs,
it could be caused by a "burr" or a buildup of film residue on the
COOLING PLATE. To check and correct for this:
a. Push the leading edge of a film over the COOLING PLATE.
b. If the film "sticks" in anyway, clean the COOLING PLATE with an
ALCOHOL WIPE.
c. If necessary, replace the COOLING PLATE.
DICOM Message
FILM JAM
"Film Jam - Error Codes 26-326 or 26-543" on
Severity
1-Alert
2G0947 - 31JAN08