Kodak DryView 5800 Service Manual page 161

Laser imager
Hide thumbs Also See for DryView 5800:
Table of Contents

Advertisement

Condition Code: 01-005
Web Portal Message
None
Condition Code: 01-007
Web Portal Message
None
2G0947 - 31JAN08
Log Message
Error stopping subsystem xxxx
Description
An error occurred while attempting to stop subsystem xxxx.
System Impact
The subsystem in question may still be running, in which case full shutdown
will not be achievable.
User Action
None
FE Action
None, this is a log notification only.
Log Message
Configuration and Control
Subsystem cannot communicate
with subsystem xxxx.
Description
An unknown error has occurred while communicating with subsystem xxxx.
System Impact
Configuration and Control will attempt to re-establish communications with the
subsystem in question. If communications is lost, the subsystem will be
shutdown by force and restarted.
User Action
None
FE Action
None, this is a log notification only.
DICOM Message
NORMAL
1-Alert
DICOM Message
NORMAL
2-Warning
Diagnostics
Severity
Severity
5-13