Condition Code: 21-624 or 23-624
Web Portal Message
xxxxx Supply: Film Calibration
Failure
2G0947 - 31JAN08
Log Message
xxxxx Supply: Calibration
Failed
Description
This error occurs when the density readings that come from the calibration print
are not monotonic or the DENSITOMETER did not find exactly 21 steps in the
calibration wedge. The most likely cause is fogged or expired film. The error is
cleared when a FILM TRAY is re-inserted after being removed.
System Impact
One or more feed magazine(s) is unable to print. The spooler is still able to
accept and queue jobs and continue printing from any other available supply
tray.
User Action
1. Try a different film pack.
2. Shut down the IMAGER, then start the IMAGER.
3. If the error persists, call for service.
FE Action
DICOM Message
CALIBRATION
1-Alert
ERR
Diagnostics
Severity
5-75