2G0947 - 31JAN08
Check:
FILM SEPARATED SENSOR
PICKUP HOME SENSOR
FILM CONTACT SENSOR
SUCTION CUPS
PICKUP DRIVE MOTOR
VACUUM PUMP
VACUUM VALVE
3. Install a new PICKUP AY (see Adjustments and Replacements,
"PICKUP AY - UPPER" on page 3-76
page
3-77).
If an existing problem:
Install a new SENSOR (see Adjustments
and Replacements,
"SENSOR, FILM
SEPARATED - UPPER (S3U) and
LOWER (S3L)" on page
Install a new SENSOR (see Adjustments
and Replacements,
"SENSOR, PICKUP
HOME - UPPER (S1U) and LOWER
(S1L)" on page
3-81).
Install a new SENSOR (see Adjustments
and Replacements,
"SENSOR, FILM
CONTACT - UPPER (S2U) and LOWER
(S2L)" on page
3-82).
Install new CUPS (see Adjustments and
Replacements,
"SUCTION CUP" on
page
3-79).
Install a new MOTOR (see Adjustments
and Replacements,
"MOTOR, PICKUP
DRIVE - UPPER (M3U) and LOWER
(M3L)" on page
3-78).
Install a new PUMP (see Adjustments and
Replacements,
"VACUUM PUMP (M4)"
on page
3-74).
Install a new VALVE (see Adjustments and
Replacements,
"VALVE, SOLENOID -
UPPER (M5U) and LOWER (M5L)" on
page
3-80).
or
"PICKUP AY - LOWER" on
Diagnostics
3-83).