Condition Code: 21-631 Or - Kodak DryView 5800 Service Manual

Laser imager
Hide thumbs Also See for DryView 5800:
Table of Contents

Advertisement

Diagnostics

Condition Code: 21-631 or 23-631

Web Portal Message
xxxxx Supply: Film Calibration
Failure - Dmin Outside Target
5-76
Log Message
xxxxx Supply: Calibration
failed - Dmin Not Met.
Description
The error occurs when the calibration sheet does not have a step density at or
below the required Dmin value. The most likely causes are expired film or
incorrect PROCESSOR temperature. The error is cleared when a new
calibration is initiated or when the FILM TRAY is removed. Any FILM TRAY
using these calibration results will automatically go into AIQC-Manual Mode
and will continue printing but with reduced range in achievable densities. A new
calibration will be attempted when this FILM TRAY is re-inserted.
System Impact
None - the system will function normally.
User Action
1. The minimum density of the film is too high. Calibration results for this
film are outside the normal range, but will still be used.
2. If you do not wish to use these calibration results, perform a new
calibration on this film or try another film pack.
FE Action
1. Check for "fogged" film. If necessary, replace the film.
2. Use the SERVICE TOOL to check the PROCESSOR temperature and set
point for correct values. If necessary, adjust the temperature (see
Adjustments and Replacements,
on page
3-1).
DICOM Message
CALIBRATION
1-Alert
ERR
"PROCESSOR - DRUM Temperature"
2G0947 - 31JAN08
Severity