Kodak DryView 5800 Service Manual page 247

Laser imager
Hide thumbs Also See for DryView 5800:
Table of Contents

Advertisement

Condition Code: 29-924
Web Portal Message
Densitometer: Internal
Diagnostic Failure
Condition Code: 29-925
Web Portal Message
Densitometer: Internal
Diagnostic Failure
2G0947 - 31JAN08
Log Message
Densitometer offset reading out
of range with light source off
Description
During a calibration print and during DENSITOMETER diagnostics, a reading
from the DENSITY SENSOR (S10) is taken with the light source turned off.
The error is indicated if the MICROCONTROLLER reports the reading out of
range. The error clears only when power is cycled.
System Impact
The printer is unable to print films. The spooler is still able to accept and queue
jobs but new films will not be initiated. Films downstream of the affected
subsystem will be completed. All other films will be reprinted.
User Action
1. Shut down the IMAGER, then start the IMAGER.
2. If the error persists, call for service.
FE Action
Log Message
Densitometer reference reading
out of range with light source on
Description
During a calibration print and during DENSITOMETER diagnostics, a reading
from the DENSITY SENSOR S10) is taken with the light source turned on. This
error occurs when the reading is out of range. (Note: Notification 29-945 is
generated when this reading is nearing the range limit.) The error clears only
when power is cycled.
System Impact
The printer is unable to print films. The spooler is still able to accept and queue
jobs but new films will not be initiated. Films downstream of the affected
subsystem will be completed. All other films will be reprinted.
User Action
1. Shut down the IMAGER, then start the IMAGER.
2. If the error persists, call for service.
FE Action
DICOM Message
ELEC DOWN
0-Fatal
DICOM Message
ELEC DOWN
0-Fatal
Diagnostics
Severity
Severity
5-99

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents