SUCTION CUP

Prerequisites:
To Remove:
To Install:
2G0947 - 31JAN08
Remove the
"PICKUP AY - UPPER"
IMPORTANT
:
This procedure is used for all PICKUP AY SUCTION CUPS.
SUCTION CUP
1. Remove each SUCTION CUP as follows:
a. Insert a small flat-blade screwdriver into the slot at the center of each
SUCTION CUP and turn counterclockwise to remove the SUCTION
CUP and HUB.
b. Remove the HUB from the SUCTION CUP.
1. Reverse the steps in the removal procedure.
Adjustments and Replacements
or
"PICKUP AY -
HUB
LOWER".
3-79