Diagnostics
Condition Code: 06-411
Web Portal Message
MIM Core: Image Data Error
Condition Code: 06-412
Web Portal Message
None
5-32
Log Message
(see Description)
Description
The page builder component has encountered the following image data error:
width and/or height is less than zero,
bits beyond the pixel depth are not zeroed.
System Impact
The current page will not be processed because it could not be rendered due to
an internal error.
User Action
1. Delete jobs in the "Unprintable jobs" queue.
2. Resend the print job from the image source.
3. If the error persists, call for service.
FE Action
Log Message
Render abort requested.
Description
The page builder component has been asked to halt a rendering that is currently
in progress.
System Impact
The current page will not be processed because rendering was aborted.
User Action
None
FE Action
DICOM Message
NORMAL
1-Alert
DICOM Message
NORMAL
1-Alert
2G0947 - 31JAN08
Severity
Severity