Diagnostics
Condition Code: 27-931
Web Portal Message
Optics: Internal
Communications Failure
Condition Code: 27-932
Web Portal Message
None
5-92
Log Message
MCS failed to communicate
with Optics (BPM) module after
multiple attempts
Description
This error occurs when the Machine Control System (MCS) tries to
communicate with the MICROCONTROLLER but receives no response to 5
attempts. The error clears only when power is cycled.
System Impact
The printer is unable to print films. The spooler is still able to accept and queue
jobs but new films will not be initiated. Films downstream of the affected
subsystem will be completed. All other films will be reprinted.
User Action
1. Shut down the IMAGER, then start the IMAGER.
2. If the error persists, call for service.
FE Action
Log Message
MCS failed to communicate
with Optics (BPM)
Description
This notification occurs when the Machine Control System (MCS) tries to
communicate with the MICROCONTROLLER but receives no response. (Note:
Error 27-931 is generated when this problem is detected on the 5th successive
failed communication attempt.)
System Impact
None - the system will function normally.
User Action
None
FE Action
None, this is a log notification only.
DICOM Message
ELEC DOWN
0-Fatal
DICOM Message
NORMAL
3-Notify
2G0947 - 31JAN08
Severity
Severity