Diagnostics
Condition Code: 27-608
Web Portal Message
None
Condition Code: 27-609
Web Portal Message
None
5-88
Log Message
Optics calibration detected laser
dynamic range approaching
predefined limits
Description
This notification occurs when the laser dynamic range has been adjusted and its
value is within 10% of the xxxxx or lower limit of the adjustment range.
System Impact
None - the system will function normally.
User Action
None
FE Action
None, this is a log notification only.
Log Message
Optics calibration detected
Optics Maximum Power
approaching limit
Description
This notification occurs when the maximum LASER power measured by the
optics calibration and diagnostics is approaching the predefined valid limit.
System Impact
None - the system will function normally.
User Action
None
FE Action
None, this is a log notification only.
DICOM Message
NORMAL
3-Notify
DICOM Message
NORMAL
3-Notify
2G0947 - 31JAN08
Severity
Severity