Kodak DryView 5800 Service Manual page 251

Laser imager
Hide thumbs Also See for DryView 5800:
Table of Contents

Advertisement

Condition Code: 29-945
Web Portal Message
None
2G0947 - 31JAN08
Log Message
Densitometer reference reading
outside of optimal range with
light source on
Description
During a calibration print and during DENSITOMETER diagnostics, a reading
from the DENSITY SENSOR (S10) is taken with the light source turned on.
This notification occurs when the reading is outside of the optimal range, i.e.
approaching the range limit. (Note: Error 29-925 is generated when this reading
is over the range limit.)
System Impact
None - the system will function normally.
User Action
None
FE Action
None, this is a log notification only.
DICOM Message
NORMAL
3-Notify
Diagnostics
Severity
5-103