Condition Code: 20-209
Web Portal Message
Laser Imager Opened During
Self Test
2G0947 - 31JAN08
Log Message
Hood or Cover Opened During
Self Test
Description
This error occurs during Machine Control System (MCS) diagnostics when the
IMAGER determines that the TOP COVER or a PANEL was opened during the
execution of Self Test. The error clears only when power is cycled.
System Impact
The printer is unable to print films. The spooler is still able to accept jobs but
new films will not be initiated. Films in progress will not be completed.
User Action
1. Close the open TOP COVER or PANEL.
2. Shut down the IMAGER, then start the IMAGER.
3. If the error persists, call for service.
FE Action
1. Check that the following are closed and secured:
•
TOP COVER
•
FRONT and BACK PANELS
2. De-energize then energize the IMAGER.
DICOM Message
PRINTER DOWN
0-Fatal
Diagnostics
Severity
5-47