Condition Code: 21-632 or 23-632
Web Portal Message
xxxxx Supply: Film Calibration
Failure - Dmax Outside Target
2G0947 - 31JAN08
Log Message
xxxxx Supply: Calibration
failed - Dmax Not Met.
Description
The error occurs when the calibration sheet does not have a step density at or
above the required Dmax value. The most likely causes are expired film or
incorrect PROCESSOR temperature. The error is cleared when a new
calibration is initiated or when the FILM TRAY is removed. Any FILM TRAY
using these calibration results will automatically go into AIQC-Manual Mode
and will continue printing but with reduced range in achievable densities. A new
calibration will be attempted when this tray is re-inserted.
System Impact
None - the system will function normally.
User Action
1. The maximum density of the film is too low. Calibration results for this
film are outside the normal range, but will still be used.
2. If you do not wish to use these calibration results, perform a new
calibration on this film or try another film pack.
FE Action
1. Use the SERVICE TOOL to check the PROCESSOR temperature and set
point for correct values. If necessary, adjust the temperature (see
Adjustments and Replacements,
on page
3-1).
DICOM Message
CALIBRATION
1-Alert
ERR
"PROCESSOR - DRUM Temperature"
Diagnostics
Severity
5-77

Hide quick links: