Condition Code: 04-110
Web Portal Message
None
2G0947 - 31JAN08
Log Message
The association request was
rejected.
Description
An association request was rejected for one or more of the following reasons:
The Service Class User (SCU) did not specify an implementation UID.
The SCU did not specify an application context name.
The protocol version is a two-byte field; bit ZERO (0) must be ONE (1).
The SCU did not specify any presentation context items to be negotiated.
None of the proposed presentation context items were successfully
negotiated. This may happen, for example, when a DICOM Store SCU
tries to communicate with a DICOM Print Service Class Provider (SCP).
System Impact
The SCU will not be able to open an association, and therefore will not be able
to deliver a print job.
User Action
None
FE Action
None, this is a log notification only.
DICOM Message
NORMAL
1-Alert
Diagnostics
Severity
5-17