Adjustments And Replacements; Adjustments; Processor - Drum Temperature - Kodak DryView 5800 Service Manual

Laser imager
Hide thumbs Also See for DryView 5800:
Table of Contents

Advertisement

Section 3: Adjustments and Replacements

Adjustments

PROCESSOR - DRUM Temperature

Adjustment Specification
Purpose:
Setting Temperature:
Specification:
Special Tools:
Prerequisites
To Check
2G0947 - 31JAN08
To set the correct temperature for the DRUM.
This procedure is required after replacing the PROCESSOR AY,
PROCESSOR DRUM, or the DRE BOARD, or if you expect that DRUM
temperatures are causing image quality problems.
The temperature at the PROCESSOR DRUM must be between 123.8° -
124.2° C (254.84° - 256.56° F).
TEMPERATURE METER with PROBE TL5574
BLOCK for PROBE, KIT SP7F3477
LAPTOP COMPUTER with Microsoft INTERNET EXPLORER 5.5
SERVICE TOOL
CROSSOVER CABLE TL5576
Note:
A TEMPERATURE METER and PROBE with a BAR TYPE
ELEMENT must be used for this procedure. PROBES with
CIRCULAR TYPE ELEMENTS will not provide correct
readings. The TEMPERATURE METER and PROBE must be
calibrated together at least once per year. If the PROBE breaks, a
new PROBE and the existing METER must be calibrated again.
The TEMPERATURE METER must be at room temperature
during this procedure. If the METER has been in a hot or cold
vehicle, allow it to adjust to room temperature before using.
None
1. Energize the IMAGER and allow it to warm to operating
temperature.
3-1

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents