Kodak DryView 5800 Service Manual page 187

Laser imager
Hide thumbs Also See for DryView 5800:
Table of Contents

Advertisement

Condition Code: 06-600
Web Portal Message
Condition Code: 10-001
Web Portal Message
MIS: Internal Software Error
2G0947 - 31JAN08
Log Message
General error in delivery
subsystem.
Description
A general, non-specific error has occurred in the Delivery Subsystem.
System Impact
Unknown.
User Action
None
FE Action
Log Message
MIS Internal Error
Description
This represents any of a number of errors that indicate that internal software
functionality of the MCS Interface Server (MIS) Subsystem has failed.
Examples include: persistent memory failures, object creations failing, system
call failures.
System Impact
The DRE is not functional.
User Action
1. Shut down the IMAGER, then start the IMAGER.
2. If the error persists, call for service.
FE Action
1. Make a copy of the log with the error.
2. Send the log to Service Engineering.
3. Make a copy of the system configuration.
4. Load new system software.
5. Restore the configuration.
DICOM Message
NORMAL
1-Alert
DICOM Message
PRINTER DOWN
0-Fatal
Diagnostics
Severity
Severity
5-39