Condition Code: 29-927
Web Portal Message
Densitometer: Internal
Diagnostic Failure
2G0947 - 31JAN08
Log Message
Densitometer fiducial sensor
reference reading out of range
with light source on
Description
During a calibration print and during DENSITOMETER diagnostics, a reading
from the FIDUCIAL SENSOR S11) is taken with the light source turned on.
This error occurs when the reading is out of range. (Note: Notification 29-947 is
generated when this reading is nearing the range limit.) The error clears only
when power is cycled.
System Impact
The printer is unable to print films. The spooler is still able to accept and queue
jobs but new films will not be initiated. Films downstream of the affected
subsystem will be completed. All other films will be reprinted.
User Action
1. Shut down the IMAGER, then start the IMAGER.
2. If the error persists, call for service.
FE Action
DICOM Message
ELEC DOWN
0-Fatal
Diagnostics
Severity
5-101