Condition Code: 06-440
Web Portal Message
None
Condition Code: 06-500
Web Portal Message
None
2G0947 - 31JAN08
Log Message
sendPage():
myMPageBuilder.RenderPage()
call has high bits masked out.
Description
The 12-bit image data had pixels that were greater than 12 bits.
System Impact
None. The system will automatically compensate for the problem.
User Action
None
FE Action
Log Message
(see Description)
Description
One of the following internal errors has occurred:
Binding error
Staging error
Printing error
Exposure error
Unknown internal error
System Impact
The job may not be delivered due to an internal error communicating with the
MCS Interface Server (MIS).
User Action
None
FE Action
None, this is a log notification only.