Adjustments and Replacements
To Install:
3-36
To remove the GASKET on the INLET SHIELD:
GASKET
1. Remove:
"PROCESSOR COVER"
2 SCREWS
INLET SHIELD
2. Pull off the old GASKET.
1. Use ALCOHOL WIPES to clean the adhesive from the GASKETS.
2. Remove the paper backing from the new GASKETS.
3. Set the GASKETS in place.
4. Press down firmly all around the GASKETS.
5. Reverse the steps in the removal procedure.
2 SCREWS
INLET SHIELD
2G0947 - 31JAN08