Kodak DryView 5800 Service Manual page 252

Laser imager
Hide thumbs Also See for DryView 5800:
Table of Contents

Advertisement

Diagnostics
Condition Code: 29-947
Web Portal Message
None
Condition Code: 36-931
Web Portal Message
Local Panel: Internal
Communications Failure
5-104
Log Message
Densitometer fiducial sensor
reference reading outside of
optimal range with light source
on
Description
During a calibration print and during DENSITOMETER diagnostics, a reading
from the FIDUCIAL SENSOR (S11) is taken with the light source turned on.
This notification occurs when the reading is outside of the optimal range, i.e.
approaching the range limit. (Note: Error 29-927 is generated when this reading
is over the range limit.)
System Impact
None - the system will function normally.
User Action
None
FE Action
None, this is a log notification only.
Log Message
MCS failed to communicate
with Local Panel after multiple
attempts
Description
This error occurs when the Machine Control System (MCS) tries to
communicate with the MICROCONTROLLER but receives no response to 5
attempts. The error clears only when power is cycled.
System Impact
The LOCAL PANEL User Interface will not display correct data.
User Action
1. Shut down the IMAGER, then start the IMAGER.
2. If the error persists, call for service.
FE Action
DICOM Message
NORMAL
3-Notify
DICOM Message
ELEC DOWN
0-Fatal
2G0947 - 31JAN08