Pm Intervals; Em Call Checklist - Kodak DryView 5800 Service Manual

Laser imager
Hide thumbs Also See for DryView 5800:
Table of Contents

Advertisement

Preventive Maintenance

PM Intervals

EM Call Checklist

4-2
IMPORTANT
:
A Preventive Maintenance (PM) call must be made every 20,000 film cycles.
To consistently produce high quality images, the IMAGER must receive
periodic routine maintenance.
Perform PM procedures:
Every EM (Emergency Maintenance Call)
Every 20,000 cycles the LOCAL PANEL displays a message prompting
the user to schedule preventive maintenance. The procedures listed in the
20,000 cycle checklist should be performed as soon as possible after the
20,000 cycle PM prompt.
If the number of prints since the last PM is more than 18,500, use the 20,000
Cycle PM Procedure instead of the EM Call Checklist.
1. Repair the print problem that caused the EM Call.
2. Open:
TOP COVER
PROCESSOR COVER
3. Check the PROCESSOR DRUM for damage.
If the PROCESSOR DRUM is damaged, replace it (see Adjustments And
Replacements,
"PROCESSOR DRUM" on page
CAUTION
:
ISOPROPYL ALCOHOL is highly flammable. Follow the
manufacturer's instructions for safe use and handling.
4. Clean the PROCESSOR COVER ROLLERS. Use ALCOHOL WIPES to
wipe the PROCESSOR COVER ROLLERS clean.
NOTE:
After cleaning, place the used wipes in a BAG (78-9998-2868-0).
TIE-WRAP the BAG and dispose of it properly.
5. If you installed a new PROCESSOR DRUM, calibrate the PROCESSOR
DRUM temperature (see Adjustments And Replacements,
- DRUM Temperature" on page
6. Complete the PM procedures.
3-27).
3-1).
2G0947 - 31JAN08
"PROCESSOR

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents