Kodak DryView 5800 Service Manual page 263

Laser imager
Hide thumbs Also See for DryView 5800:
Table of Contents

Advertisement

Item
LED6*
LIFT_HOME LOWER - When lit, lower FILM SUPPLY PICKUP BAR is at home (up).
LED7*
FILM_SAVER UPPER - When lit, upper FILM SAVER is not present.
LED8*
FILM_SAVER LOWER - When lit, lower FILM SAVER is not present.
LED9, LED10
FILM_OUT UPPER, FILM_OUT LOWER - not used, never lit.
LED11*
FILM_CONTACT UPPER - When lit, the upper PICKUP is not touching film in the upper
FILM SUPPLY.
LED12*
FILM_CONTACT LOWER - When lit, the lower PICKUP is not touching film in the
lower FILM SUPPLY.
LED13
LASER_ILOCK - When lit, this LED indicates LASER power is enabled. It does not mean
the LASER is energized. LASER power is enabled only when:
LED14*
FILM_SEP UPPER - When lit, there is no film separation (the SUCTION CUPS are not
lifting a sheet of film) from the upper FILM SUPPLY.
LED15*
FILM_SEP LOWER - When lit, there is no film separation (the SUCTION CUPS are not
lifting a sheet of film) from the lower FILM SUPPLY.
LED16
DOOR_ILOCK - When lit, the 24 V motor drive voltage (Hazard 24 V) is enabled. This
voltage is enabled when: .
LED17, LED18
TRAY_LOCK UPPER, TRAY_LOCK LOWER - not used, never lit.
LED19*
FILM_PATH2 - not used, always lit.
LED20*
NIPOC_LIMIT, UPPER - Indicates that the NIP ROLLER is moving from open to closed
or from closed to open.
LED21
NIPOC_LIMIT, LOWER - not used, never lit.
LED22
FILM_PATH3 - not used, never lit.
LED23*
CASS_IN UPPER - When lit, there is no FILM TRAY in the upper FILM SUPPLY.
LED24*
CASS_IN LOWER - When lit, there is no FILM TRAY in the lower FILM SUPPLY.
LED25
FILM_PATH4 - not used, never lit.
LED26, LED27
BAG UPPER, BAG LOWER - not used, never lit.
HAZARD +24V
Test Point for Hazard 24 volts (MOTOR supply voltage). This voltage is enabled when:
2G0947 - 31JAN08
Table 5-7: FPM LEDs and Test Points (Cont'd)
The 3 INTERLOCK SWITCHES are closed and,
The SERVICE SWITCH is in the Normal position.
The 3 INTERLOCK SWITCHES are closed or,
The SERVICE SWITCH is in the Service position.
The 3 INTERLOCK SWITCHES are closed or,
The SERVICE SWITCH is in the Service position.
Description
Diagnostics
5-115