Kodak DryView 5800 Service Manual page 264

Laser imager
Hide thumbs Also See for DryView 5800:
Table of Contents

Advertisement

Diagnostics
Item
+24V TEST
Test Point for Digital 24 volts.
+12V TEST
Test Point for Digital 12 volts.
+V5 TEST
Test Point for Digital 5 volts.
+V3_3 TEST
Test Point for Digital 3.3 volts.
GND1, GND2,
Test Points for Logic Ground.
GND3, GND4
JP2 (Jumper pins)
Jumper across both pins overrides the MICROPROCESSOR SENSOR control and enables
all FILM SUPPLY SENSORS.
S1
RESET SWITCH - Pressing this SWITCH causes a Checksum and Memory test to run for
the software on the FPM.
S2
DEFAULTS SWITCH.
* Indicates FILM SUPPLY SENSOR LEDs - - - Each of these LEDs is controlled by one of the FILM SUPPLY
SENSORS. LED ON = Related SENSOR blocked.
5-116
Table 5-7: FPM LEDs and Test Points (Cont'd)
Description
2G0947 - 31JAN08