Diagnostics
Severity Levels
Severity
0-Fatal
A failure occurred and the IMAGER will
have to be restarted.
1-Alert
A failure occurred, but it should be
temporary and correctable by the user.
2-Warning
An event occurred that should notify the
user.
3-Notify
An event occurred, but was automatically
recovered.
4-Auxiliary
Status information.
5-10
Table 5-5: Subsystem IDs (Cont'd)
Subsys ID
Subsystem/Component Name
MCS:
20
- General MCS System
21
- Upper FILM SUPPLY
22
- N/A For The 5800 IMAGER
23
- Lower FILM SUPPLY
24
- N/A For The 5800 IMAGER
25
- RF Tag
26
- FILM TRANSPORT
27
- OPTICS
28
- PROCESSOR
29
- DENSITOMETER
30 - 35
- N/A For The 5800 IMAGER
36
- LOCAL PANEL User Interface
The following table describes the severity levels associated with condition codes
for the IMAGER.
Table 5-6: Severity Levels
Description
Recovery
The IMAGER must be restarted.
The IMAGER should be able to continue after
the problem is corrected. If the failure is FILM
SUPPLY specific, the IMAGER should be able
to continue operating using the other FILM
SUPPLY.
No recovery needed.
A notice is written to the log. The IMAGER
should be able to continue automatically.
Does not impact IMAGER operation.
Error ID
20-xxx
21-xxx
23-xxx
25-xxx
26-xxx
27-xxx
28-xxx
29-xxx
36-xxx