6) When the auto-calibration function is finished, the detector displays the Main Menu
screen, which should now look something like this. (Your screen will look different if
IXR is enabled.)
Product Name
Notice that, in our example, the auto-calibration function has learned the following basic
parameters for our uncontaminated product.
Detect level
Phase angle
What the Main Menu Screen Is Telling You
The Main Menu screen is now telling you that, once the detector is monitoring your
product stream, any contaminated product that exceeds the detect level of 202, will be
rejected. And, because the phase angle is 1.46 degrees, the detector is monitoring a wet
product.
Page 92
= 202
= 1.46
Thermo Scientific APEX User's Guide