Thermo Scientific APEX 100 User Manual page 175

Apex series
Table of Contents

Advertisement

9) Press the left-navigation button to select the "Accept" check mark.
Accept
10) Press the Go button. (The detector pastes in the product parameters and closes the
confirmation screen.)
11) Press the Back button repeatedly to return to the Main Menu, which should now
look like this.
Product Name
Notice that the initial setting for detect level and phase angle for the "MILK—P.M. SHIFT"
(shown in step 6, where the settings were 40 and 90, respectively) have changed, and
now match those for the "MILK—A.M. SHIFT" (25 and 2, respectively, as shown in the
screen in step 2 above).
Summary
All "MILK—A.M. SHIFT" parameters have now been copied to the "MILK—P.M. SHIFT."
Clearly, by using copy and paste, you have made a substantial saving in time and effort
in comparison to the alternative—re-entering the product parameters, re-running the
calibrations, and so on.
Thermo Scientific APEX User's Guide
Page 175