cycle, the reject will not be confirmed, and the flow of product can be stopped and an
alarm triggered notifying you of the problem.
Keying In a Value for the Reject 1 Complete
To enable the "reject-1 complete" function for gravity-feed, pipeline, and pharmaceutical
applications, do as follows.
1) Make sure that your "reject-1 confirm" and "reject-1 complete" inputs have been
enabled and the sensors hardwired to the appropriate input on the detector's wiring
board.
2) Make sure the "reject-1 confirm" function is highlighted (as shown above).
3) Press the Go button and an input screen appears.
Reject 1 Complete Time
4) Key in an appropriate time window (in seconds and hundredths of a second) in which
the reject must occur. Typically the calibration function will set the appropriate value
and no manual adjustment is necessary.
5) Press the Go button to save your setting and exit the menu.
Page 140
Thermo Scientific APEX User's Guide