Thermo Scientific APEX 100 User Manual page 411

Apex series
Table of Contents

Advertisement

Reject Duration Time
This is the time that the gate remains closed
to divert contaminated product, and is usually
set to one second.
1) Navigate to the screen shown in the
"Reject-Delay Time" section.
2) Highlight the "Reject 1 Duration" function.
Reject 1 Duration
3) Press the Go button and an input screen
appears.
4) Key in the appropriate value (in seconds
and hundredths of a second) for the
reject-duration time.
Calibrate X and R Noise
Thresholds
See the instructions on page 5.
Full-Product Calibration
See the instructions on page 5.
Pipeline Applications
Listed below are the most commonly used
functions for setting up a typical pipeline
application.
Reject-Delay Time
This function is usually set to 0.00 seconds,
and tells the diverter valve to close
immediately when contaminants are detected.
If your diverter valve is a long way from the
search head, you may need to set a longer
delay time.
1) Make sure the Main Menu is displayed.
2) Navigate to the rejects menu.
3) Press the Go button.
(continued...)
Page 6
4) On page 1 of the rejects menu, highlight
the "Reject 1 Distance/Delay" function.
Reject 1 Distance/Delay
5) Press the Go button and an input screen
appears.
6) Key in an appropriate value (in seconds
and hundredths of a second) for the
reject-delay time.
Reject Duration Time
This is the time that the diverter valve
remains closed to divert contaminated
product to the waste tank, and is usually set
to 3.00 seconds.
1) Navigate to the screen shown in the
"Reject-Delay Time" section.
2) Highlight the "Reject 1 Duration" function.
Reject 1 Duration
3) Press the Go button and an input screen
appears.
4) Key in the appropriate value (in seconds
and hundredths of a second) for the
reject-duration time.
Calibrate X and R Noise
Thresholds
See the instructions on page 5.
Full-Product Calibration
See the instructions on page 5.
Thermo Scientific APEX Quick-Start Guide
REC-F-58E—Part #085555

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Apex 300Apex 500

Table of Contents