How the Detector Calculates the AuditCheck Signal Time-Out
When the AuditCheck signal time-out is set to zero, the detector automatically calculates
the value using the following formula.
Signal Time-Out (ms) =
Where: Search-Head Width (SHW) is in mm, Overhang Distance (OD)
is in mm, Filter Delay is in ms, and Product Speed is in m/min.
Keying In a Value for Signal Time-Out
To key in a value for the signal time-out, do the following.
1) Navigate to page 4 of the AuditCheck menu.
AuditCheck - Setup
2) Navigate to the signal time-out function and make sure it is highlighted.
AuditCheck Signal Timeout
3) Press the Go button and an input screen appears.
AuditCheck Signal Timeout
4) Key in your signal time-out in milliseconds.
5) Press the Go button to save your setting and exit the function.
Thermo Scientific APEX User's Guide
60 * 1.75 * [SHW + (2 * OD)]
Product Speed
+ Filter Delay
Page 213