at least 30 cm.) Maintaining the appropriate pack gap is critical to the success of the
calibration procedure.
To auto-calibrate the IXR function, do the following.
1) Make sure the detector's Main Menu is displayed.
Product Name
2) Press the Go button and page 1 of the "Product Calibrate and Name" menu appears.
Product Calibrate and Name
3) Navigate to the "Full Product Calibration" menu and make sure it is highlighted.
Full Product Calibration
4) Make sure the conveyor is running at the speed you will use during normal
production runs, and press the Go button to start the IXR calibration. The following
screen appears.
Performing Full Learn
Thermo Scientific APEX User's Guide