To key in a value for pulse length, do the following.
1) Navigate to page 4 of the AuditCheck menu.
AuditCheck - Setup
2) Navigate to the pulse-length function and make sure it is highlighted.
AuditCheck Pulse Length
3) Press the Go button and an input screen appears.
AuditCheck Pulse Length
4) Key in your pulse length (in milliseconds).
5) Press the Go button to save your setting and exit the function.

Setting the AuditCheck Signal Time-Out

When the AuditCheck solenoid fires, a time-out commences during which the detector
waits for the AuditCheck signal. Thus, the AuditCheck time-out signal is the length of
time it takes the shuttle to travel the length of the AuditCheck tube, plus the speed-filter
delay. However, if the internally generated signal time-out pulse—for some reason—
does not give the exact result you want, you can manually enter a value.
Page 212
Thermo Scientific APEX User's Guide