Calibrating the Reject-2 Confirm Time

In order to calibrate the reject 2 confirm time, that is, have the APEX set the appropriate
time for you, you must do the following.
Turn off your reject-1 device by going to the Outputs menu and setting the
logically assigned value to X (off), as shown below. For detailed instructions on
doing this, see the "Output" section of the manual.
So, change the reject-1 output from this:
Output 1 Function
To this:
Output 1 Function
Set your reject-2 device to reject contaminants during the calibration process.
(Described below.)
Pass a contaminated package, so the reject-2 device can reject it, allowing the
APEX to set the reject-2 confirm time. (Described below.)
Turn your reject-1 device back on, which is a critical step! (Described below.)
Calibration Procedure
To calibrate the reject-2 confirm time, do the following.
1) Navigate to page 5 of the rejects menu, and highlight the "Reject 2 Contaminant"
menu.
Reject 2 Contaminant
(continued...)
Thermo Scientific APEX User's Guide
Page 147