To start a manual AuditCheck, do the following.
1) Make sure the manual AuditCheck menu is highlighted, as shown below.
Manual Test
2) Press the Go button to start the AuditCheck.

Setting Limits for Peak Signal and Phase Angle

These functions have already been described in detail for the Quality Test function in the
Quality Test section, starting on page 196—so please refer to the appropriate sections in
the Quality Test write-up for detailed instructions, as the Peak Signal and Phase Angle
Parameters are very similar for AuditCheck and the Quality Test function.
AuditCheck – Warning, Alarm, Fault levels

Setting the AuditCheck Overhang Distance

This function allows the filter delay to be accurately tracked, when an AuditCheck test is
performed.
1) Navigate to page 4 of the AuditCheck menu.
AuditCheck - Setup
Page 210
Thermo Scientific APEX User's Guide