2) Press the Go button and page 1 of the WAF menu appears.
Warnings, Alarms & Faults
3) Navigate to the bin-full event menu.
Bin Full event
4) The bin-full error message currently triggers a fault notification. To change the bin-
full error message to an alarm, press the Go button repeatedly to scroll through the
error-message options (described above). When you reach the single alarm icon (no
other icons are present in the menu), press the Back button to save your setting and
exit the menu.
5) Now, in our example, when the reject bin is full, the detector lights the small red
lamp you connected to Output 5. This alarm notifies you the reject bin is full.
Page 216