Establishing Basic Product Parameters

Before putting your detector into service, you must have it learn how to identify your
uncontaminated product. This is done using the detector's full-product calibration
function, which sets the following basic parameters for your product.
Phase-angle setting
Detect level
Doing a Full-Product Calibration
To do a full-product calibration, follow the instructions below.
1) Make sure product is flowing through the pipe.
2) Highlight your product name in the Main Menu. (Your screen will look different if IXR
is enabled.)
Product Name
3) Press the Go button and page 1 of the product menu appears.
Product Calibrate and Name
(continued...)