Setting the Reject Device Used for the QA Test

This function sets the reject device used during the QA test. The default setting is to
assign Reject 1 to the Quality Test function.
1) Navigate to the reject device menu and make sure it is highlighted.
Reject Device
Note: The X indicates that, currently, no reject device is assigned to the Quality Test
function.
2) Press the Go button repeatedly to scroll through the reject-device options, which are
as follows.
Icon
Meaning
No reject device is assigned
Assign reject-device number 1 (Reject 1)
Assign reject-device number 2 (Reject 2)
Note: For your reject device to work, it must be connected to the detector's wiring board
and be assigned to a specific output (Output 1–6). For more details, see pages 290
and 232.
3) When the option you want is shown in the screen, press the Back button to save
your selection and exit the menu. (We have chosen to use our reject-1 device for our
quality testing.)
Reject Device
Page 202