Keying In the Signal Duration for the Reject Device

Before starting, make sure you know the signal-duration time (expressed in seconds and
hundredths of a second) needed to activate your main (Reject 1) device.
1) If you are not already on page 1 of the rejects menu, follow steps 1–2 in the "Keying
In Distance to the Reject Device" section to get there.
2) Navigate to the reject-device signal-duration function.
Reject 1 Duration
The screen above shows that, currently, the reject-1 device remains active for 1.50
seconds.
3) Press the Go button and the signal-duration input screen appears. Use the navigation
buttons to key in your signal duration—in seconds and hundredths of a second. (In
our salted butter example, we have entered our signal-duration time of 2.02
seconds.)
Reject 1 Duration
4) Press the Go button to save your settings and exit the input screen.
5) Press the Back button repeatedly to return to the Main Menu.
Page 46
Thermo Scientific APEX User's Guide