Frequency And Gain; Calibrating The Detect Level - Thermo Scientific APEX 100 User Manual

Apex series
Table of Contents

Advertisement

Frequency and Gain—Page 2
1) From the Main Menu, use the navigation buttons to select the frequency and gain
menu.

Frequency and Gain

2) Press the Go button and page 1 of the frequency and gain menu appears. Press the
down-navigation button to get to page 2.
Detection level and Noise Threshold

Calibrating the Detect Level

This function allows you to calibrate only the detect level (rather than having the detect
level set by the Main Menu's auto-calibration function, which simultaneously calibrates
the phase angle, detect level, and AuditCheck settings).
Calibrate Detect Level
Calibrating Only the Detect Level
To calibrate only the detect level (typically used in a conveyor application), do the
following.
1) Turn the conveyor on and have some samples of uncontaminated (metal-free)
product ready to pass through the search head.
(continued...)