3) Navigate (as described in the above sections) to the speed-encoder function, and
make sure the speed-encoder function is highlighted, as shown below.
Speed Encoder
4) Press the Go button, and the monitor screen (below) shows a bar graph of the
progress of the calibration process. During the calibration, the detector's speed-
encoder function automatically matches the pulses from the conveyor's speed
encoder to the speed you entered in the product-speed function.
Calibrating Encoder Frequency
5) When the calibration is completed, the detector displays the following screen. (The
value shown in your screen, however, may be different.)
Speed Encoder
6) Press the Back button repeatedly to return to the Main Menu.
Page 126
Thermo Scientific APEX User's Guide