Keying In the Distance to the Reject Device

Before starting, make sure you know the distance from the downstream edge of the
search head to the center of your main (Reject 1) device.
1) From the Main Menu navigate to the reject-totals menu.
Reject Totals
2) Press the Go button and page 1 of the rejects menu appears.
Reject 1 Distance and Duration
3) Navigate to the "Reject 1 Distance/Delay" function
Reject 1 Distance/Delay
Stop and look very carefully at the highlighted number in the screen above, and note
that the current setting is 0.300 meters—as shown by the m after the number. (In other
words, the reject distance is not expressed in seconds, which for some applications,
would be a valid setting.)
a. If the number in the "Reject 1 Distance/Delay" menu is shown as a time unit (as
indicated by an s after the number), you should now change it to a distance unit.
To do so, press the right-navigation button and highlight the change-units
function, as shown below.
Reject 1 Distance/Delay
Units
Page 44
Thermo Scientific APEX User's Guide