5) Press the Go button. The detector resets the product parameters—including deleting
the name "MILK"—and returns you to the Main Menu.
Product Name
Notice that all parameters for Product 003 have now been reset to their factory defaults.
The detect level, for example, has been reset to 40 and phase angle to 90.

Resetting the NVRAM Parameters to the Default Settings

This function resets virtually everything in the detector to its factory-default settings.
Only supervisors can enter this function. However, do not enter this function unless you
are firmly resolved to delete all your product and other settings and start afresh. This
function clears everything from the detector's memory—except the model of APEX (500,
300, or 100) and type of application (conveyor application, gravity-feed, pipeline or
pharmaceutical) you are using. You have been warned! Proceed with caution.
Reset NVRAM to Defaults
Resetting the NVRAM
1) Make sure the NVRAM function is highlighted, as shown above.
2) Press the Go button and the keyboard screen appears.
3) Type in your supervisor password. If you have not assigned a unique password to
yourself, you should use the default password "SUPE" here (that is, enter only the
four upper-case letters, SUPE).
4) In the keyboard screen, highlight the exit-and-save key.
(continued...)
Page 186
Thermo Scientific APEX User's Guide