The Six Product Maximum Has Been Reached
If you attempt to set up more than six products (the maximum allowed by the IXR
function), the following information box appears.
Solution: Delete one of the previously set up products.
Signal Saturation Has Occurred
If signal saturation occurs while you are calibrating the IXR function, the following
warning box appears.
Solution: Go to the "Frequency and Gain" menu and adjust these settings to a lower
level. If the problem persists, please contact Thermo Fisher Scientific—as described in
the "Contacting Thermo Fisher Scientific" section at the end of the manual.
Page 392
Thermo Scientific APEX User's Guide