8) Press the Back button repeatedly to return to the Main Menu. Please note that, now
the IXR function is active, additional displays have been added to the Main Menu, as
shown below. (Please see the "Understanding the IXR Display" section on page 388
for more information about these additional displays.)
Main Menu

Calibrating the IXR Function

There are two basic steps to calibrating the IXR function, as follows.

Learn the X and R noise thresholds.

Learn the detect levels (for both the X and R signals).
In addition, if you are going to use the IXR function for long production runs lasting
several hours or days, you should make sure the IXR tracking function is set to "On"
(the default value).
Learn the X and R Noise Thresholds
Please note it is very important that, while calibrating the X and R noise thresholds, you
run the conveyor at exactly the same speed you will use during your normal production
runs. Thus, if you decide at a later date, to change the speed of the conveyor during
your normal production runs, you must recalibrate the X and R noise thresholds.
To calibrate the X and R noise thresholds, do the following.
1) Make sure the Main Menu is displayed.
Product Name
(continued...)