Elzone II 5390 Confirm
Procedure
If possible, prepare your sample when you are ready to start the analysis. If you prepare
your sample in advance, be sure to place a lint-free cover over the beaker to prevent the
entry of air-borne particles.
1.
Place approximately 0.2 gram of sample into a 25-mL beaker.
If your sample tends to agglomerate, add a few drops of surfactant. Mix until it
resembles paste.
2.
Add 20 mL of electrolyte to the beaker.
3.
Place the beaker in an ultrasonic bath for approximately 1-2 minutes.
If sample particles are fragile, do not use the ultrasonic bath; gently stir with a
spatula instead.
4.
Clean the external glassware (orifice tube, electrode, and stirrer):
a. Fill a squeeze bottle with clean, filtered electrolyte or water.
b. Hold a beaker under the glassware and, using the squeeze bottle, thoroughly rinse
each piece of glassware.
5.
Place the analysis beaker on the sample stand; position the electrode next to the
orifice tube so that it will be submerged in the beaker.
3-34
Preparing for an Analysis