Elzone II 5390 Confirm
Calibration
4-18
Displays the date and time for the type of calibration
selected. Click
Change
calibration; the Calibration Type dialog is displayed.
Reference material measurement
Uses data from a reference material measurement.
Mass integration measurement
Before the analysis, displays Measured; after analysis, it
will display the date and time of the analysis that was
used for the measurement.
Entered calibration factor
Enter the desired calibration factor. The following table
provides the nominal Kd values for Elzone orifice tubes.
Nominal Kd values for Elzone Orifice Tubes
Orifice Size
19
30
48
76
95
150
190
300
380
Calibration copied from sample
Uses calibration data from a completed sample file that
used the same orifice tube; the
enabled allowing you to choose the file.
to choose a different type of
Kd
33
42
60
93
120
185
250
448
582
push button is
Browse