Appendix A
6130-
The calculation could not be performed.
Cause:
Action A:
Action B:
6131-
The peak height has not been initialized, and so cannot be used in the fit.
Cause:
Action A:
Action B:
6132-
Calibration failed: No SRM selected.
Cause:
Action:
539-42822-01 (Rev A) - Nov 2017
You requested a calculated peak height on the orifice tube calibration
review screen and the number of reference materials is greater than the
number of peaks (see Message 6136).
Select the peak height manually.
Make sure the correct sample is in the system, the correct SRM's are
selected, and the proper analysis conditions (for the current sample) are
selected; then try again.
You requested that a peak be included in the fit but the peak height had
not been assigned.
Select and assign the peak height manually.
Make sure the correct sample is in the system, the correct SRM's are
selected, and the proper analysis conditions (for the current sample) are
selected; then try again.
No SRMs were selected on the Orifice Tube calibration review screen.
Select one or more SRMs to be included.
Elzone II 5390 Confirm
A-39